Máy đo nước đa chỉ tiêu 2 kênh có máy in (PH/ ORP/ ISE)
Model: TruLab pH/ISE 1320P
SKU: 1GA340PY
Hãng sản xuất: YSI – Mỹ
Xuất xứ: Mỹ
1. Tính năng kỹ thuật:
- The YSI TruLab line includes the 1110 (single channel), 1310 and 1310P (single channel) and 1320 and 1320P (dual channel) instruments providing easy-to-use and calibrate instruments ideal for the laboratory.
- Measure pH, mV (ORP), ISE and temperature accurately in the laboratory with the YSI pH 1320 or 1320P with built-in printer. The TruLab pH/ISE 1320 benchtop meter is the ideal laboratory instrument for reliable measurements with automatic documentation for GLP compliance.
- The 1320 has internal memory. Stored data can be sent to a PC via USB or to an optional built-in printer available on the 1320P.
- Simultaneous dual measurement of pH, mV (ORP) or ISE and temperature (2-channel benchtop meter)
- AutoRead function improves repeatability
- Easy-to-read backlit display with text menu for easy operation for pH, mV (ORP), ISE and temperature
- Fast data transfer via USB in .csv or ASCII format or the optional integrated printer
- Easy calibration with adjustable calibration timer Automatic or manual temperature compensation
- 1, 2, 3, 4 or 5 point calibration for pH
- 1, 2, 3, 4, 5, 6 or 7 point calibration for ISE, also non- linear
- BNC connection for pH, mV and ISEs; banana connection for temperature probes
- Continuous Monitoring Control (CMC) function for monitoring the ideal measuring range for pH measurements based on the last calibration
- GLP/AQA compliant
- Wide selection of electrodes - pH, ORP, ISE and temperature probes
- PC software, USB cable, and electrode stand included
- Parameters:
§ pH, mV (ORP/Redox)
§ Temperature
§ ISEs: Ammonia, Ammonium, Bromide, Cadmium, Calcium, Chloride, Copper, Cyanide, Fluoride, Iodide, Lead, Nitrate, Potassium, Silver/Sulfide, Sodium
2. Thông số kỹ thuật:
Benchtop
|
Yes
|
Buffer Recognition
|
Yes
|
Certifications
|
CE, cETLus, IP43, electrical protective class III
|
Display
|
Backlit graphic display
|
Equipment used with
|
TruLab analog pH/mV/ISE/temp probes
|
GLP Compliance
|
Yes
|
Graphic Display
|
Backlit graphic display
|
Logging Capabilities
|
Yes
|
Measurement Method / Channels
|
2 Channel
|
Measurement Range
|
Ranges Resolutions Accuracy (+1 digit) pH -2.0 to 20.0; -2.00 to 20.00; -2.000 to 19.999 0.1; 0.01; 0.001 ±0.1; ±0.01; ±0.005 (temp of test sample 15 to 35°C; within a range of ±2 pH of the calibration point) mV -1200 to 1200.0; -2500 to 2500 0.1; 1.0 ±0.3; ±1.0 (temp of test sample 15 to 35°C) ISE (mg/l, mg/kg, mmol/l, ppm, %) (µmol/l) 0.000 to 9.999; 10.00 to 99.99; 100.0 to 999.9; 1000 to 999999 0.000 to 9.999; 10.00 to 99.99; 100.0 to 999.9; 1000 to 9999 0.001; 0.01; 0.1; 1 Temperature -5 to 105°C (23.0 to 221°F) 0.1 ±0.1
|
Memory
|
5,000 data sets auto logging, 500 data sets manual logging
|
Operating pH Range
|
-2.000 to 19.999, -2.0 to 20.0, -2.00 to 20.00
|
Operating Temperature
|
23.0 to 221°F, -5 to 105°C
|
Parameters Measured
|
pH, ORP (mV), Ammonia, Ammonium, Bromide, Cadmium, Calcium, Chloride, Copper, Cyanide, Fluoride, Iodide, Lead, Nitrate, Potassium, Silver/Sulfide, Sodium, Temperature
|
Power
|
4 alkaline AA batteries fully charged or included universal power supply, approximately 150 hours with illumination or 1000 without
|
Sampling
|
Yes
|
Storage Temperature
|
-25 to 65°C, -13 to 149°F
|
Unit of Measure
|
Sample addition, ISE Measurements: Blank value correction, increment methods: Known addition (single), Known subtraction, Known addition (double), Sample subtraction
|
User Calibratable
|
Yes
|
User Replaceable
|
Yes
|