Multi-parameter analysers
Model: C5010 •C5020
Manufacture: CONSORT – Belgium
Origin: Belgium
Reference at: http://www.consort.be/
1. Overview:
pH:
- Multi-point (1...3) calibration with up to three buffers out of eleven pre-programmed pH buffers.
- Reads pH with 0.01 pH resolution.
mV:
- Features mV calibration for accurate ORP measurements.
- Reads potentials with 1 mV resolution.
Conductivity
- Use a 1 cm-1electrode to measure from 0.1 µS/cm to 100 mS/cm.
- Use a 0.1 cm-1 electrode to measure from 0.01 µS/cm to 10 mS/cm. (C5020 only)
- Use a 10 cm-1electrode to measure from 1 µS/cm to 1000 mS/cm. (C5020 only)
- Automatically selects correct range and frequency.
- Selectable reference temperature: 20° or 25°C.
- One-point calibration.
Dissolved oxygen:
- Operates with a galvanic oxygen electrode requiring no polarisation time and no zero calibration.
- Reads dissolved oxygen with 0.01 mg/l or 0.1% resolution.
- Rapid air calibration.
Temperature:
- Reads temperatures with 0.1°C resolution.
- Manual or automatic temperature compensation.
- Calibrates temperature probe for quality measurements.
Inputs:
- One common input for pH and mV.
- One common input for conductivity and dissolved oxygen.
- One input for a Pt1000 automatic temperature probe.
Display:
- Bright LCD screenfor better readability.
- A white backlight automatically illuminates when operated on the mains.
- Stability indicator prompts the user when readings should be taken.
- The interactive LCD screen provides step by step instructions in the language of your choice (English, Dutch, French, German).
- Shows a GLP report on the LCD screen.
Data-logging: (C5020 only)
- Storage memory for 300 values including temperature.
Cabinet:
- Robust dust and splash-proof cabinet.
Special features:
- Three year warranty.
- Mains and rechargeable battery operation with programmable auto-matic switch-off.
- Optional 12 V car adaptor.
Pre-programmed standards:
- pH buffers: 1.68, 2.00, 4.00, 4.01, 6.87, 7.00, 9.18, 9.21, 10.01, 12.00, 12.45 (at 25°C).
- Conductivity: 1413 µS/cm, 12.88 mS/cm, 111.8 mS/cm (at 25°C).
2. Technical Data:
pH:
- Range: 0...14 pH
- Resolution: 0.01 pH
- Accuracy: 0.2% ± 1 digit
- Calibration: 1...3 points
- Buffers: 11 pre-programmed
- Temperature: compensation 0...100°C
- Slope: 80...120%
mV:
- Range: ±1000 mV
- Resolution: 1 mV
- Accuracy: 0.2% ± 1 digit
- Calibration: 1 point
rH2:
- Range: 0...42 rH2
- Resolution: 0.1 rH2
CONDUCTIVITY:
- Range (cc dependent):
§ 0...100 mS/cm (C5010)
§ 0...1000 mS/cm (C5020)
- Resolution (cc dependent):
- 0 ...µS/cm (C5010)
- 0...0.01 µS/cm (C5020)
- Accuracy: 1% f.s. of range
- Calibration: 1 point
- Standards: 3 pre-programmed
- Cell constant (cc):
§ 1 cm-1 ±30% (C5010)
§ 0.1/1/10 cm-1±30% (C5020)
- Temperature compensation: 0...100°C
- Reference temperature: 20° or 25°C
- Temperature coefficient: natural waters (EN27888)
SALINITY: (C5020 only)
- Range: 0.0...70.0
- Reference temperature: 15°C
TDS: (C5020 only)
- Range: 0...100 g/l
- Resolution: 0.1 mg/l
DISSOLVED OXYGEN:
- Range: 0...20 mg/l (0...200%)
- Resolution: 0.01 mg/l (0.1%)
- Accuracy: 1% ± 1 digit
- Calibration: 1 point
- Temperature compensation: 0...50°C
- Salinity compensation: 0...40
- Air pressure compensation: 800...1200 hPa
TEMPERATURE:
- Range: 0...100°C 0...100°C
- Resolution: 0.1°C 0.1°C
- Accuracy: 0.5°C 0.5°C
- Calibration: 1 point 1 point
INPUTS:
- pH/mV: BNC, 1012 Ω BNC, 1012Ω
- Conductivity/Dissolved oxygen: BNC
- Temperature: 2 banana, for Pt1000 2 banana, for Pt1000
MEMORY: (C5020 only)
- Data sets: 300 300
DISPLAY:
- LCD: 122x32 pixels 122x32 pixels
- White backlight
AMBIENT CONDITIONS:
- Temperature: 0...40°C 0...40°C
- Humidity: 0...95%, non condensing 0...95%, non condensing
POWER SUPPLY:
- Mains: 100...240 VAC, 50/60 Hz 100...240 VAC, 50/60 Hz
- Low voltage: 9...15 VDC 9...15 VDC
- Batteries: 4x1.2 V, NiMH 4x1.2 V, NiMH
DIMENSIONS (WxDxH): 10x20x4 cm 10x20x4 cm
WEIGHT Meter: 350 g 350 g
3. Order Info:
Item
|
Code
|
Description
|
01
|
C5010
|
pH/conductivity/DO meter
|
02
|
C5020
|
pH/conductivity/DO meter
|
03
|
C50x0P
|
Meter kit for pH: meter + pH/ATC electrode SP10T + 2x50 ml buffers (pH 4 and 7) + 50 ml electrolyte (3M KCl) + carrying case
|
04
|
C50x0K
|
Meter kit for conductivity: meter + conductivity/ATC electrode SK10T + 50 ml conductivity standard (0.01 M KCl) + carrying case
|
05
|
C50x0Z
|
Meter kit for oxygen: meter + dissolved oxygen electrode SZ10T + carrying case
|
06
|
C50x0T
|
Meter kit complete: meter + pH/ATC electrode SP10T + conductivity/ATC electrode SK10T + dissolved oxygen electrode SZ10T + 2x50 ml buffers (pH 4 and 7) + 50 ml electrolyte (3M KCl) + 50 ml conductivity standard (0.01 M KCl) + carrying case
|
07
|
C50x0X
|
Meter kit without electrodes: meter + 2x50 ml buffers (pH 4 and 7) + 50 ml electrolyte (3M KCl) + 50 ml conductivity standard (0.01 M KCl) + carrying case
|
08
|
A4049
|
Car adaptor, 12 V (optional)
|
Không có nhận xét nào: