Thứ Năm, 1 tháng 6, 2017

Multi-parameter analysers Model: C3040 •C3041 Manufacture: CONSORT – Belgium Origin: Belgium

Multi-parameter analysers
Model: C3040 •C3041
Manufacture: CONSORT – Belgium
Origin: Belgium
Reference at: http://www.consort.be/


1.      Overview:
pH:
-          Multi-point (1...5) calibration for more linearity.
-          Selectable resolution from 0.001 pH to 0.1 pH.
-          Automatic calibration with any of eleven pre-programmed and five user specified pH buffers.
-          Accepts pH electrodes with any zero point (Eo) between ±999 mV.
mV:
-          Features mV calibration for accurate ORP measurements.
-          Selectable resolution from 0.1 mV to 1 mV.
Ion:
-          Direct concentration measurement.
-          Multi-point (2...5) calibration and an additional blank correction for measuring low concentrations.
Conductivity:
-          Multi-point (1...3) calibration for more linearity.
-          An electrode with a typical cell constant of 1 cm-1 permits to measure from 0.01 µS/cm to 200 mS/cm in five ranges.
-          An electrode with a typical cell constant of 0.1 cm-1 permits to measure from 0.001 µS/cm to 20 mS/cm in five ranges.
-          An electrode with a typical cell constant of 10 cm-1permits to measure from 0.1 µS/cm to 2000 mS/cm in five ranges.
-          Automatically selects correct range and frequency.
-          Selectable reference temperature: 20° or 25°C.
-          Automatic calibration with any of three preprogrammed and three user specified standard solutions.
-          Allows to lock the initial conductivity range to avoid non-linear titration curves.
-          Accurate low conductivity measurements by eliminating the capacitive component of the electrode and its cable (avoid the use of long cables!).
Dissolved oxygen:
-          Operates with a galvanic dissolved oxygen electrode requiring no polarisation time and no zero calibration.
-          Selectable resolution from 0.01 mg/l (0.1%) to 0.1 mg/l (1%).
-          Automatic air pressure compensation 600-1300 hPa.
Temperature:
-          Reads temperatures with 0.1°C resolution.
-          Manual or automatic temperature compensation (O2: 0...50°C).
-          Calibrates temperature probe for quality measurements.
Inputs:
-          Two inputs for pH, mV, Ion, dissolved oxygen or conductivity + corresponding temperature and reference inputs.
-          Four extra inputs for pH, mV, Ion or dissolved oxygen + corresponding temperature and reference inputs.
Outputs: Two versions available:
-          C3040: with USB communication port (galvanically isolated) and RS232 interface.
-          C3041: with Ethernet communication port and RS232 interface.
Display:
-          A large bright LCD screen with white backlight enables to view all
-          channels individually or simultaneously.
-          Stability indicator prompts the user when readings should be taken.
-          Hold function allows to freeze the display for convenient reading or recording.
-          The interactive LCD screen provides step by step instructions in the language of your choice (English, Dutch, French, German).
-          Real-time clock displays time and date.
-          Shows a GLP report on the LCD screen.
Data-logging:
-          Up to 12000 data sets can be stored manually or at a programmable interval.
-          Allows to mix data from all ranges in the same table.
Cabinet:
-          Robust dust and splash-proof cabinet.
-          An optional wall mounting kit allows to fix the meter to any wall making more space available on the desk.
Special features:
-          Two-way communication with a computer using USB, Ethernet or RS232.
-          Can be programmed to continue automatically with the measure-ments or data-logging after a power failure.
-          Password protection prevents any unauthorised modification of the instrument's settings.
-          No electrical interference between pH/ORP/Ion and conductivity electrodes in the same solution.
-          Optional 12 V car adaptor.
-          Three year warranty.
GLP:
-          All procedures for a "Good Laboratory Practice" are on board.
Pre-programmed standards:
-          pH buffers: 1.68, 2.00, 4.00, 4.01, 6.87, 7.00, 9.18, 9.21, 10.01, 12.00, 12.45 (at 25°C).
-          Conductivity: 1413 µS/cm, 12.88 mS/cm, 111.8 mS/cm (at 25°C).

2.      Technical Data:
pH:
-          Range: -2...+16 pH
-          Resolution: 0.001 pH
-          Accuracy: 0.1% ± 1 digit
-          Calibration: 1...5 points
-          Buffers: 11 pre-programmed; 5 user specified
-          Temperature compensation: -5...+105°C
-          ISO-pH: 6...8 pH
-          Slope:  80...120%
-          Zero point (Eo): ±999 mV
mV:
-          Range: ±2000 mV
-          Resolution: 0.1 mV
-          Accuracy: 0.1% ± 1 digit
-          Calibration: 1 point
ION:
-          Range: 0.01 ng/l...100 g/l
-          Resolution: 3 digits
-          Accuracy: 0.5% ± 1 digit
-          Calibration: 2...5 points + blank
CONDUCTIVITY:
-          Range (cc dependent): 0...2000 mS/cm
-          Resolution (cc dependent): 0.001 µS/cm
-          Accuracy: 0.5% f.s. of range
-          Calibration: 1...3 points
-          Standards: 0.01/0.1/1 M KCl; 3 user specified
-          Cell constant (cc):  0.07...13 cm-1
-          Temperature compensation: -5...+105°C
-          Reference temperature: 20° or 25°C
-          Temperature coefficient: natural waters (EN27888)
-          Range lock
-          Capacitive compensation
RESISTIVITY:
-          Range: 0...200 MΩ.cm
-          Resolution: 1 Ω.cm
SALINITY:
-          Range: 0.0...70.0
-          Reference temperature: 15°C
TDS:
-          Range: 0...100 g/l
-          Resolution: 0.01 mg/l
DISSOLVED OXYGEN:
-          Range: 0...60 mg/l (0...600%)
-          Resolution: 0.01 mg/l (0.1%)
-          Accuracy: 1% ± 1 digit
-          Calibration: 1 point
-          Temperature compensation: 0...50°C
-          Salinity compensation: 0...40
-          Air pressure compensation: 600...1300 hPa
TEMPERATURE:
-          Range: -5...+105°C
-          Resolution: 0.1°C
-          Accuracy: 0.1°C
-          Calibration: 1 point
AIR PRESSURE:
-          Range: 600...1300 hPa
-          Calibration: 1 point
CHANNELS:
-           Measurement: 6 (conductivity: 2)
-          Temperature: 6
INPUTS:
-          Measurement: 6 BNC, 1012
-          Temperature: 6x2 banana, for Pt1000
CALIBRATION:
-          Reminder: 0...999 h
-          GLP
DISPLAY:
-           LCD: 240x64 pixels
-          White backlight
-          Hold function
-          Selectable resolution
-          Real time clock
COMMUNICATION:
-           Interface with computer: USB or Ethernet
-          RS232, baud rate: 1200...115200 b/s
-          Printer
DATA-LOGGING:
-           Data sets: 12000 + °C/date/time
-          Modes all
-          Manual or timed
-          Interval: 1...9999 s
SECURITY:
-           Identification number
-          Password protection
AMBIENT CONDITIONS:
-           Temperature: 0...40°C
-          Humidity: 0...95%, non condensing
POWER SUPPLY:
-           Mains 100...240 VAC, 50/60 Hz
-          Low voltage 9...15 VDC
DIMENSIONS (WxDxH): 26x18x9 cm
WEIGHT: Meter 1 kg

3.      Order Info:

Item
Code
Description
01
C3040
pH/Ion/conductivity/DO meter (USB version) + USB cable
02
C3041
pH/Ion/conductivity/DO meter (Ethernet version) + UTP cable
03
SH300
Flexible electrode holder (optional)
04
A4800
Wall mounting kit (optional)
05
A4049
Car adaptor, 12 V (optional)

Không có nhận xét nào:

Đăng nhận xét